Prerequisite:
- Check that the lever of the venting tool is in the closed position.
To empty the gas from the cylinder:
- On the cylinder valve, unscrew the plug on the pressure gauge port.
- Screw the venting tool with the M12x1 connector on the pressure gauge port of the cylinder valve.
- Tighten to the stop using a 19 mm wrench.
- Operate the lever to open the ball valve.
- When the cylinder is emptied, operate the lever again to close the ball valve.
- Unscrew the venting tool.
⇨ The non-return valve in the pressure gauge port closes automatically.
Gas under pressure: risk of injury
After operating the lever, the gas released through the hole may cause injury.
Suffocation hazard
Cylinders must be vented outdoors.